"Vodafone": +380996603539 "Kyivstar": +380677331007, +380686027409

Indicator of electrical machines windings defects IDO-07


IDO-07 – is a portable instrument designed for monitoring windings of electrical machines.
IDO-07 – provides detection of:

  • three-phase windings for the presence of turn-to-turn short circuits, phase loss and incorrect connection of the phases;
  • coils of windings laid in slots, for the presence of turn-to-turn short circuits;
  • insulation states of the windings relative to the machine body and between the windings.

IDO-07 allows you to detect defects in a three-phase winding of a machine with a voltage of up to 1000 V in the assembly, as well as to identify defects in a distributed winding of a stator or rotor of a disassembled machine with a voltage of up to 1000 V (except for windings with equalizing connections).

Operating principle:

– when checking the three-phase winding for the presence of turn-to-turn short circuits, phase loss and the correct connection of the phases, the currents of the two phases of the winding are compared when an alternating voltage up to 10 V is applied to them with a frequency up to 10 kHz. If there are defects, the phase currents will be different;
– when checking the coil of the winding put in the slots for the presence of turn-to-turn short circuits, a pulse EMF of 1V / turn is induced in it. If there are short – circuited turns in the coil, the magnetic induction pulse of the field generated by the short-circuit current flowing through them is registered;
– when checking the insulation state of the windings relative to the machine body and between the windings, a DC voltage is applied to the winding, the insulation resistance is determined and the latter is compared with the threshold value (0.5 MOhm).

Technical specifications:

1) testing parameters:
– when checking the three-phase winding for the presence of turn-to-turn short circuits, phase loss and the correct connection of the phases asymmetry coefficient of the phase currents, Kn
– when checking the coils of the windings laid in the slots for the presence of turn-to-turn short circuits current in the checked coil
– when checking the insulation condition of the windings relative to the machine body and between the windings insulation resistance, Ri
2) controlled range Kn, % 0÷99
3) controlled range Ri, MOhm 0÷500
4) Kn value when closing one turn in phase, %, not less 10
5) output DC voltage when measuring Ri, V 1000±100
6) indication LED
7) power supply stand-alone or from an external power supply
8) supply voltage, V 5
9) power consumed, W, no more 4
10) overall dimensions, mm 205×805×50
11) weight, kg 0.4

The warranty period – is 24 months from the date of sale.